Table 8.3, Output registers – RLE LD5100 User Manual
Page 52
www.rletech.com
52
970.484.6510
8
Modbus Communication
Table 8.3 Output Registers
Register
Name
Description
Units
Range
40001
Leak Threshold
Trip current for leak alarm
Amps
0-65535
40002
Contamination
Threshold
Trip current for contamination alarm
Amps
0-65535
40003
Re-Alarm
Re-Alarm delay
Minutes
0-65535
40004
Latched Alarm
Latching Alarms 0=No
1=Yes
0-65535
40005
Silence Alarm
Set to 1 to silence audible alarm
1=Silence
0-65535
40006
Reset Alarm
Set to 1 to reset alarms
1=Reset
Alarm
0-65535
40007
Spare
0-65535
40008
Spare
0-65535
40009
Spare
0-65535
40010
Month
Clock
1-12
0-65535
40011
Day
Clock
1-31
0-65535
40012
Year
Clock
00-99
0-65535
40013
Hour
Clock
0-23
0-65535
40014
Minutes
Clock
0-59
0-65535
40015
Seconds
Clock
0-59
0-65535
40016
Seconds
Leak Alarm Delay
20-3600
0-65535
40017
Seconds
Contamination Alarm Delay
20-3600
0-65535