9 Indicating transmitter DTR specifications

85

9.4 Specifications

9.4.1 Indicating transmitter specifications

Display:

320x240 pixel graphical LCD with LED
backlight

Keypad:

18 membrane keys

Current output:

Two independent current sources, 4-20 mA,
max. load 1000 Ohm, galvanic isolation
1500 VDC or AC (peak), hold function during
prism wash

Power:

AC input 100-240 VAC

±10%,

50/60 Hz,30 VA, optional 24 VDC;
Overvoltage Category II, Pollution Degree 2
(DTR), Pollution Degree 3 (sensor)

Alarms/Wash relays:

Two built-in signal relays, max. 250 V/3 A

Input switches:

Four switch inputs.

Current outputs:

Two current outputs configurable
independently to indicate process
concentration or temperature of either
sensor.

Sensor connectivity, DTR:

One or two sensors can be connected to the
transmitter. Sensors independent of each
other: own parameter sets and usable in
different applications.

Sensor connectivity, STR:

Only one sensor can be connected to the
transmitter.
Used with the Intrinsically safe sensor
PR-21-S-...-IA.

Transmitter protection class:

Enclosure IP66, Nema 4X

Installation:

Indoor/outdoor, relative humidity 80% max

Indicating transmitter weight:

4.5 kg (10 lbs)

9.4.2 Interconnecting cable specifications

Cable:

IEC 61158-2 compliant two-wire cable:
two signal wires and shield
copper area 0.8 mm

2

(18 AWG)

cable resistance 24 Ohm/km (per wire)
cable attenuation 3.0 dB/km @ 28 kHz

Cable length:

Standard 10 m (33 ft), max. total length
200 m (660 ft)