12 Sensor verification

117

Figure 12.3

Verification display

Important:

Clean and dry the prism and the sample holder very carefully be-

tween the R.I. liquids. Use a suitable solvent, e.g., ethanol.

To check that the standard liquid is properly wetting the prism, it is possible to
press soft key

OPT. IMAGE

. The optical image should show like in Figure 12.4. For

more information on the optical image, please see section 5.4.1.

Figure 12.4

Optical image

A verification data collection method is implemented in the DTR. The instrument
measures each verification data point ten times and uses the average of these mea-
surements. Measuring each verification liquid takes approximately ten seconds,
during which the measurement progress display (Figure 12.5) is shown. Please
wait until the verification step 2 display reappears before proceeding to next ver-
ification liquid.

Figure 12.5

Verification in process

Pressing

COMPLETE

in the verification display finishes the verification procedure and

calls up the verification results.