K-Patents PR-21-S User Manual
Page 123

12 Sensor verification
117
Figure 12.3
Verification display
Important:
Clean and dry the prism and the sample holder very carefully be-
tween the R.I. liquids. Use a suitable solvent, e.g., ethanol.
To check that the standard liquid is properly wetting the prism, it is possible to
press soft key
OPT. IMAGE
. The optical image should show like in Figure 12.4. For
more information on the optical image, please see section 5.4.1.
Figure 12.4
Optical image
A verification data collection method is implemented in the DTR. The instrument
measures each verification data point ten times and uses the average of these mea-
surements. Measuring each verification liquid takes approximately ten seconds,
during which the measurement progress display (Figure 12.5) is shown. Please
wait until the verification step 2 display reappears before proceeding to next ver-
ification liquid.
Figure 12.5
Verification in process
Pressing
COMPLETE
in the verification display finishes the verification procedure and
calls up the verification results.