6 Configuration and calibration

53

Note: NAMUR is an international association of users of automation in process
industries. The association recommendation NE 43 promotes a standardization
of the signal level for failure information. The goal of NE 43 is to set a basis for
proactively using transmitter failure signals in process control strategies. Using
these failure signals, instrument faults are separated from process measurements.

NAMUR NE 43 uses the 3.8 to 20.5 mA signal range for measurement information,
with ≥21 mA or ≤3.6 mA to indicate diagnostic failures (see figure 6.11). With
that information, it is easier to detect a failure condition on a refractometer, for
example, it clearly tells you whether you have an empty pipe or a failed instrument.

0

3.6

3.8

4

20

20.5

21

mA

failure

OK

Measurement Data

failure

Figure 6.11

Default mA output values

6 MANUAL SET

allows you to set different output values to check the output signal.

Press the

BACK

key to return normal output function.

Note: If you want to ’turn off’ the mA output, select

NOT DEFINED

in the Source menu.