3 wash – K-Patents PR-21-S User Manual
Page 84

78
PR-21-S instruction manual
8.3 Wash
8.3.1 Message
EXTERNAL HOLD
The concentration measurement is on HOLD due to an external switch closure. For
explanation, see Section 6.4, “Configuring input switches”.
8.3.2 Messages
PRECONDITIONING
,
WASH
,
RECOVERING
−
PRECONDITIONING
: An optional preconditioning relay is closed, see Section 6.7
“Configuring prism wash”.
−
WASH
: The internal wash relay is closed, see Section 6.7 “Configuring prism
wash”.
−
RECOVERING
: The concentration measurement is on HOLD during a preset time.
8.3.3 Message
PRISM WASH FAILURE
No dip of n
D
value during prism wash. The accepted size of the dip is set as the
WASH CHECK
function, Section 6.7, “Configuring prism wash”. See also Section 5.1.3
“Testing prism wash”.
8.3.4 Message
EXTERNAL WASH STOP
Tells that wash action is prevented because an
EXTERNAL WASH STOP
switch is closed,
see Section 6.4 “Configuring input switches”.
−
NO SAMPLE
(Section 8.2.6) indicates empty pipe
8.3.5 Message
LOW TEMP WASH STOP
Tells that wash action is prevented because of
LOW TEMP
: low process temperature
indicates empty pipe. To set the limit, see Section 6.7 “Configuring prism wash”.
8.3.6 Message
NO SAMPLE/WASH STOP
Tells that wash action is prevented because of
NO SAMPLE
: the process pipe is empty
and the prism is clean.