12 Sensor verification

115

12 Sensor verification

A company maintaining quality system according to ISO 9000 quality standards
must have defined procedures for controlling and calibrating its measuring equip-
ment. Such procedures are needed to demonstrate the conformance of the final
product to specified requirements. The company should

− Identify the required accuracy and select appropriate equipment for measure-

ments.

− Establish calibration procedures including a check method and acceptance cri-

teria.

− calibrate the equipment at prescribed intervals against certified equipment

having a known valid relationship to nationally recognized standards. In cases
where no such standards exist, the basis used for calibration must be docu-
mented.

K-Patents verifies the calibration of all delivered instruments according to a pro-
cedure similar to the one described Section 12.1. K-Patents quality system is
ISO 9001 certified by Det Norske Veritas.

12.1 Refractive index n

D

verification

Before starting the verifications procedures, make sure that you have a K-Patents
PR-21-S sample holder at hand. Also check the condition of your standard refrac-
tive index liquids. You’ll also need a cleaning solution (ethanol) to clean the sensor
prism and the sample holder.

Sample holder (Figure 12.1) is kept in place by a rubber link.

Figure 12.1

Sample holder - Part number PR-5002