5 startup and use, 1 startup – K-Patents PR-21-S User Manual
Page 41

5 Startup and use
35
5 Startup and use
5.1 Startup
5.1.1 Initial check
1. Check the wiring, Section 3.3, “Electrical connections”.
2. Connect the power. The Power indicator light (Figure 3.4) and the screen
should light up within a few seconds.
3. The Main display should come up on the display, Figure 5.1.
Main display for two sensors
Main display for single sensor,
concentration only
Main display for single sensor,
concentration and temperature
Main display for single sensor,
concentration and bar graph
Figure 5.1
Main display alternatives
4. In case the display shows a row of dashes, there is no corresponding sensor
(for example in Figure 5.1, upper left, there’s no sensor A, only sensor B is con-
nected). The diagnostic message is for that sensor
NO SENSOR
.
5. Check the serial number of the sensor at the upper right corner in the display.
6. For a connected sensor, the diagnostic message at start-up should be
NORMAL OP-
ERATION
or, if the process pipe is empty,
NO SAMPLE
. Otherwise, see Section 8.4,
“Diagnostic messages table”.