beautypg.com

K-Patents PR-21-S User Manual

Page 62

background image

56

PR-21-S instruction manual

LAB%

Sample concentration determined by the user

From DTR display:

(see Figure 6.13)

CALC

Calculated concentration value

T

Process temperature measurement in Centigrade

nD

Actual refractive index n

D

CONC

Measurement in concentration units, the large size number

In addition to the calibration data, write down the Indicating transmitter serial
number, the sensor serial number and the sensor position, i.e. whether it is in-
stalled as Sensor A or as Sensor B.

Accurate calibration is only achieved if the sample is taken correctly. Pay special
attention to following details:

− The sampling valve and the refractometer should be installed close to each

other in the process.

!

Warning! Wear protective clothing appropriate to your process when operating
the sampling valve and handling the sample.

− Run the sample before starting to collect data points to avoid sampling old

process liquid that has remained in the sampling valve.

− Read the values

CALC

,

T

(emp),

nD

and

CONC

in the DTR’s display at exactly the same

time with sampling.
The easiest way of doing this is to use the

FIELD SAMPLE

soft key available through

the Sensor status display (DTR program version 2.0 or newer). The value of
each sample is the average of 10 consequent measurements to increase accu-
racy and reduce possible process noise.

Press the

FIELD

SAMPLE

soft key...

...wait while

DTR measures...

(take sample for lab)

...and press

REPEAT

for next

data point or BACK to get to

the Sensor status display.

Figure 6.13

Using

FIELD SAMPLE

soft key

− Use a tight container for the sample to avoid evaporation.

Important:

Offline calibration using process liquid very seldom gives reliable

results, as problems are caused by

− low flow which makes sample to form an unrepresentative film on the prism

− sample evaporation at high temperature or undissolved solids at low temper-

ature giving deviations from laboratory determinations

− an ageing sample which is not representative

− outside light reaching the prism

Thus calibration using the process liquid should always be made inline.