beautypg.com

5 configuring refractometer system – K-Patents PR-21-S User Manual

Page 58

background image

52

PR-21-S instruction manual

6.5 Configuring refractometer system

The Indicating transmitter has two built-in 4–20 mA outputs (

mA OUTPUT 1, mA OUT-

PUT 2

), two relay contact outputs (

RELAY 1, RELAY 2

), and four switch inputs (

SWITCH 1,

SWITCH 2, SWITCH 3, SWITCH 4

). Each of these resources can be freely assigned to either

sensor A or sensor B.

6.5.1 Configuring mA outputs

For the electrical properties of the two output signals, see Section 3.3.3, “Connect-
ing the Indicating transmitter”.

− First select

5 CALIBRATION

in the Main menu and enter password if necessary. Then

select

2 OUTPUTS

in the Calibration menu. In the Outputs menu, select

5 mA OUTPUTS

.

− Select the mA output, 1 or 2, to get to the Output menu (Figure 6.10 below)

where the output can be configured.
Note: The line at the bottom of the Output menu display indicates the current
configuration of the selected mA output, e.g. in Figure 6.10 mA Output 1 has
been configured to send the concentration reading of Sensor B.

Figure 6.10

The Output menu for mA Output 1

− To change the sensor the selected output is assigned to, select

1 SENSOR

in the

Output menu.

− To change output source for the selected output, select

2 SOURCE

.

Note: Selecting

1 NOT DEFINED

’turns off’ the selected output.

− The

3 ZERO

value sets the value when the signal is 4 mA. The default zero value

is 0.00, the unit depends on the source and display unit set for the sensor in
question (and can thus be for example 0 BRIX or 0 °F).

− The

4 SPAN

sets the range, i.e. the value given when the signal is 20 mA.

Example: If your measurement unit is CONC% and you want to measure the range
15–25 CONC%, first choose concentration as mA output source. Then set the zero
value at 15 and span at 10. This means that the output signal is 4 mA at 15 CONC%
and 20 mA at 15+10=25 CONC%. To change this output to range 10–30 CONC%,
change zero to 10 and span to 20 (10+20=30).

5 DEFAULT OUTPUT

sets a mA default output value that the instrument returns to in

certain malfunction situations. The value can be set to a low or high mA value,
e.g. 3.0 mA or 22 mA. For a list of malfunctions that are affected, see section 8.4.