beautypg.com

K-Patents PR-21-S User Manual

Page 67

background image

6 Configuration and calibration

61

6.7.2 Setting prism wash parameters

To set the prism wash parameters for a given sensor, first select the sensor, then
select

5 CALIBRATION

from the Main menu, and then

4 PRISM WASH

. This menu contains

the alternatives (factory settings are given in parentheses):

1

PRECONDITION TIME

0–30 s (0 s)

2

WASH TIME

0–30 s (3 s)

3

RECOVERY TIME

0–120 s (20 s)

4

WASH INTERVAL

0–1440 min (20 min)

5

WASH CHECK MODE

(Disabled)

6

HOLD DURING WASH

(Active)

7

TEMP LIMIT ACTIVATION

8

TEMP LIMIT VALUE °C

9

EMPTY PIPE CHECK

(Active)

1

WASH nD LIMIT (on next ’page’)

The prism wash cycle: See Figure 6.16 and Section 6.7.1. The timing of the wash
cycle is controlled by the

WASH INTERVAL

,

PRECONDITION TIME

,

WASH TIME

and

RECOVERY TIME

settings. If the

WASH INTERVAL

is set to zero, the wash can be initiated only by using

the manual wash or remote wash request.

If the

PRECONDITION TIME

is zero (or there is no relay configured for preconditioning),

the preconditioning phase is skipped. If the

WASH TIME

is zero (or there is no relay

for wash), the wash functionality is completely disabled.

The recommended wash times and wash medium pressures are given in Sec-
tion 4.5, “Recommended wash pressures and times”.

Wash check: The prism wash check monitors automatically that the wash really
has an effect on the prism. In the

WASH CHECK STANDARD

mode, prism wash is accepted

if the refractive index n

D

either falls below wash n

D

limit (default 1.34) at

NORMAL

OPERATION

or

NO SAMPLE

occurs. This is the indication of a successful wash with water

or steam.
If the wash is not accepted, the diagnostic message

PRISM WASH FAILURE

(see Sec-

tion 8.4) will appear. This message is reset by a successful wash.
The

WASH CHECK AUTOMATIC WASH CUT

mode differs from the standard mode by stopping

the wash 2 seconds after the n

D

falls below the limit.

To stop the measurement for the duration of the prism wash, choose

6 HOLD

DURING WASH

and in that menu activate the hold function. The CONC reading and mA

output are held in the value they had immediately before starting the wash cycle.

To activate (or deactivate) a temperature limit, choose

7 TEMP LIMIT ACTIVATION

and

then the appropriate command in the menu.

To set a low temperature limit, choose

8 TEMP LIMIT VALUE °C

and enter the temper-

ature (in °C!) where the limit should be.

The empty pipe check prevents washing if message is

NO SAMPLE

, i.e. there’s no

process liquid in the pipe. To deactivate (or active) the empty pipe check, choose

9 EMPTY PIPE CHECK

and then the appropriate menu command.

To change the wash n

D

limit, select first

0 MORE ...

and then

1 WASH nD LIMIT

to set the

n

D

value to be used with the wash check functionality.