50
PR-21-S instruction manual
Figure 6.7
The manual set display for relays
6.4 Configuring input switches
For the electrical properties of the four input switches, see Section 3.3. To see
which switches are closed, check the Description menu, see Section 5.3. To config-
ure the switches, follow the instructions below:
1. Select Menu to get to the Main menu.
2. Select
5 CALIBRATION
from the Main menu.
3. Select
5 SWITCHES
from the Calibration menu.
4. Select the switch, 1, 2, 3 or 4, to be configured. You now get the Switch menu,
see Figure 6.8 below.
Figure 6.8
Switch menu
5. First select
1 SENSOR
to assign the chosen switch to a given sensor.
Note: The selection line will automatically go to the currently valid setting,
i.e. in Figure 6.9 next page Switch 1 has been assigned to Sensor A.