beautypg.com

K-Patents PR-21-S User Manual

Page 57

background image

6 Configuration and calibration

51

Figure 6.9

Arrival to Switch 1’s sensor selection menu, sensor A selected

6. In the Switch menu, select

2 FUNCTION

to set switch function.

Note:

The current assignment of the switch is shown at the bottom of the

Switch menu display, e.g. Figure 6.9 Switch 1 is assigned to Sensor A with func-
tion Wash stop.

1

NOT DEFINED

Factory setting

2

HOLD

When used with a built-in wash relay, this func-
tion is useful for an intermittent process: the prism
is washed when the process stops (as indicated by
contact closure). The wash is repeated when the
process restarts (if the stop lasts over 60 seconds).
The signal is on hold between washes. When used
with an external independent timer, contact closure
holds the output signal.

3

WASH STOP

Switch closure prevents wash cycle. It can be used to
prevent wash action when the process pipe is empty.
The message

WASH STOP

will appear when a wash cycle

is initiated.

4

REMOTE WASH

At switch closure the system waits for an external
wash command before initiating wash.

5

SCALE SELECT

Any chemical curve and associated field calibration
scale can be selected by switch closure. The scales
assigned to each switch independently.

6

CALIBRATION SEAL

Contact closure prevents access to calibration and
configuration ("external password"). Can be used to
seal the calibration.

7. If you chose Scale select as switch function, return to the Switch menu (if not

returned automatically) and choose

3 SCALE CHEMICAL

to enter the parameters for

the chemical curve assigned to the switch. See Section 6.6.1 for more informa-
tion on chemical curves and chemical curve parameters.

8. If necessary, the chemical curve assigned to a switch can be adjusted by field

calibration parameters. Select

4 SCALE FIELD

in the Switch menu to enter the pa-

rameters. See Section 6.6.3 for more information on field calibration and field
calibration parameters.