beautypg.com

6 calibrating the concentration measurement – K-Patents PR-21-S User Manual

Page 60

background image

54

PR-21-S instruction manual

6.6 Calibrating the concentration measurement

mA OUTPUT

FIELD CALIBRATION

CHEMICAL CURVE

n

D

CCD

TEMP

Pt-1000

Indicating

transmitter

DTR

Sensor

CONC

CALC

1.

2.

3.

4.

5.

6.

DAMPING

Figure 6.12

The six layers of

concentration calibration

The concentration calibra-
tion of the K-Patents inline
refractometer PR-21-S is or-
ganized in six layers.

1. The information from the

CCD element and the Pt-1000

temperature element. The
position of the shadow
edge (Figure 2.2, “Image
detector system”) is de-
scribed by a number called
CCD and scaled from 0–100 %.

2. The sensor calibration: The

actual refractive index n

D

is calculated from the CCD

value. The process tem-

perature is calculated from
the Pt-1000 resistance.
The sensor output is n

D

and temperature TEMP
in Centigrade. Hence, the calibrations of all PR-21-S sensors are identical,
which makes sensors interchangeable. Furthermore, the calibration of each
sensor can be verified using standard refractive index liquids, see Chapter 12.

3. The chemical curve: The Indicating transmitter DTR receives n

D

and TEMP and

calculates the concentration value according to chemical curves derived from
available chemical literature and K-Patents expertise. The result is a tempera-
ture compensated calculated concentration value CALC.

4. Field calibration: Adjustment of the calculated concentration value CALC may

be required to compensate for some process conditions or to fit the measure-
ment to the laboratory results. The Field calibration procedure, Section 6.6.3,
determines the appropriate adjustment to CALC. The adjusted concentration
is called CONC. If there is no adjustment, CALC and CONC are equal. Thus the
chemical curve is kept intact as a firm base for the calculation, the adjustment
is merely additional terms.

5. Damping: See Section 6.1.

6. Output signal: The range of the 4 to 20 mA signal is defined by its two endpoints

on the CONC scale, see Section 6.5.1.