beautypg.com

3 measurement subsystem – AMETEK 2001RP User Manual

Page 97

background image

User and Programming Manual - Rev R

2001RP

87

10.3 Measurement Subsystem

This subsystem programs the measurement capability of the RP Series AC source.

Subsystem Syntax

:MEASure

:VOLTage?

Returns the rms value of voltage

:CURRent?

Returns the rms value of current

:AMPLitude

:MAX?

Returns peak current

:RESet

Reset the peak current measurements

:CREStfactor?

Returns current crest factor

:RANGe

Selects measurement range 0 = High, 1 = Low, 2 = AUTO

:POWer

[:REAL]?

Returns real power in Watt

:APParent?

Returns apparent power in VA

PFACtor?

Returns power factor

MEASure:VOLTage?

This query returns the AC rms voltage being sourced at the output terminals. Note that the
output relay must be closed or no voltage will be present at the output terminals.

Query Syntax

MEASure:VOLTage?

Parameters

None

Examples

MEAS:VOLT?

Returned Parameters

Related Commands

OUTP 0

 1

MEASure:CURRent?

This query returns the rms value of the output AC current being sourced at the output
terminals. Note that the output relay must be closed to obtain current flow.

Query Syntax

MEASure:CURRent?

Parameters

None

Examples

MEAS:CURR?

Returned Parameters

Related Commands

OUTP 0

 1

MEAS:CURR:AMPL:MAX?