beautypg.com

AMETEK 2001RP User Manual

Page 95

background image

User and Programming Manual - Rev R

2001RP

85

CALibrate:MEASure:POWer[:FSCale]

This command affects the calibration of the power measurement at full scale.

Command Syntax

CALibrate:MEASure:POWer[:FSCale]

Parameters

(value range -127 to +127)

Examples

CAL:MEAS:POW 78

Query Syntax

CALibrate:MEASure:POWer?

Returned Parameters

(value range -127 to +127)

Related Commands

CALibrate:MEASure:CURRent

CALibrate:MEASure:POWer:ZERO

This command affects the calibration of the power measurement zero offset.

Command Syntax

CALibrate:MEASure:POWer:ZERO

Parameters

(value range -127 to +127)

Examples

CAL:MEAS:POW:ZERO -5

Query Syntax

CALibrate:MEASure:POWer:ZERO?

Returned Parameters

(value range -127 to +127)

Related Commands

CALibrate:MEASure:POWer

CALibrate:MEASure:VOLTage[:FSCale]

This command affects the calibration of the rms voltage measurement at full scale.

Command Syntax

CALibrate:MEASure:VOLTage[:FSCale]

Parameters

(value range -127 to +127)

Examples

CAL:MEAS:VOLT 56

Query Syntax

CALibrate:MEASure:VOLTage?

Returned Parameters

(value range -127 to +127)

Related Commands

CALibrate:MEASure:CURRent

CALibrate:MEASure:VOLTage:ZERO

This command affects the calibration of the rms voltage measurement zero offset.

Command Syntax

CALibrate:MEASure:VOLTage:ZERO

Parameters

(value range -127 to +127)

Examples

CAL:MEAS:VOLT:ZERO +4

Query Syntax

CALibrate:MEASure:VOLTage:ZERO?

Returned Parameters

(value range -127 to +127)

Related Commands

CALibrate:MEASure:VOLTage