2.6 Environmental
Parameter
Specification
Operating Temp:
0 degrees to + 40 degrees Celsius.
Storage Temp:
- 40 degrees to + 85 degrees Celsius.
Humidity:
Operating:
≤ 90% RH up to 40° C., non condensing
Storage:
≤ 90% RH up to 40° C, ≤75% RH up to 70° C.
Altitude
Operating:
15000 feet / 6000 meters maximum.
VA Power:
Full VA power output :0 - 6000 feet / 0 - 2000
meters.
Derate VA power 5 % per 1000 feet from 6000 to
15000 feet / 2000 to 6000 meters.
Creepage and
Clearance:
Rated for Pollution Degree 2.
Insulation:
Rated to Installation Category (Overvoltage Category) II
Vibration:
Designed to meet NSTA 1A transportation levels.
Shock:
Designed to meet NSTA 1A transportation levels.
2.7 Regulatory
Parameter
Specification
Electromagnetic
Emissions and
Immunity:
Tested to EN50081-1 and EN50082-1 European Emissions and
Immunity standards as required for the “CE” marking.
Acoustic Noise:
65 dBA maximum at 0% to 50% load, 75 dBA maximum greater
than 50% load to 100% load. Measured at one meter.
Safety:
Tested to EN61010-1 European safety standards as required for
the “CE” marking.