AMETEK 2001RP User Manual

Page 106

background image

User and Programming Manual - Rev R

96

March 2008

2001RP

OUTPut: STARt[:STATe]

The output start state command is used to enable or disable the start phase logic of the AC
source controller. If disabled (State = 0), any voltage drops with the OUTP:DROP command
or voltage changes with the VOLT command are executed immediately without regard for the
phase angle of the output sinewave.
If enable (State = 1), any voltage drops or voltage changes take place at one of four user
specified phase angles. See the OUTP:STAR:PHAS command for available phase angle
settings.

Command Syntax

OUTPut:STARt[:STATe]

Parameters

0 | 1

*RST Value

0

Examples

OUTP:STAR 1

Query Syntax

OUTP:STAR?

Returned Parameters

0 | 1

Related Commands

OUTP:DROP

OUTP:STAR:PHAS

OUTPut: STARt:PHASe

The output start phase command is used to set the start phase angle for drop or voltage
programming commands. This phase angle information is used only if the
OUTP:DROP:STATe is on (1). If not, all drops and voltage changes occur at random phase
angles.
Available phase angle values are 0°, 90°, 180° and 270°. All other values when passed as a
parameter to this command will generate a syntax error.

Command Syntax

OUTPut:STARt:PHASe

Parameters

0 | 90 | 180 | 270

*RST Value

0

Examples

OUTP:STAR:PHAS 90

Query Syntax

OUTPut?

Returned Parameters

0 | 90 | 180 | 270

Related Commands

OUTP:DROP

OUTP:STAR:STAT