4.1.1 Status Indicator Lights

Ten LED status indicators are located directly above the LCD display. These LED’s
correspond to the following conditions:

REMOTE

The REMOTE LED indicates that the unit is in remote
control mode. If the RS232C interface is used, the REMOTE
state can be enabled by the controller using the SYST:REM
command. Any time the REMOTE LED is lit, the front panel
of the RP Series unit is disabled. There is no LOCAL button
that allows the user to regain control of the front panel. The
SYST:LOC command will enable the front panel controls.
When using IEEE, the remote /local state is controlled by
the REN (Remote Enable) interface line.

FAULT

The FAULT LED indicates an overtemperature condition.
Overtemperature is usually caused by poor air flow. Check
the air flow exhaust at the rear of the unit to make sure it is
not obstructed. If the power source is configured for
constant voltage, the indicator will illuminate for loss of
sense lines or an overload condition.

OUTPUT

The Output LED indicates the status of the OUTPUT relay.
The output relay is controlled by the OUTPUT ON/OFF
button located directly below the LED. When the Output
LED is not lit, the output voltage is not present at the output
terminal regardless of the voltage setting.

RANGE - HIGH

The Range High LED indicates the selected output voltage
range. When it is illuminated it indicates the high voltage
range has been programmed. Any time the user switches
voltage ranges, the output is programmed to zero volt. The
output relay state is not changed however.

RANGE - AUTO

The Range AUTO LED indicates the 2001RP is in auto
ranging mode and will switch voltage range automatically if
the selected output voltage increments or decrements
through the low range voltage limit. (150 V

RMS

on standard

units). During a range change, the output voltage will be
programmed to zero momentarily. Some hysteresis is
provided. See Figure 4-2.

FREQ

Illuminates when the LCD display shows the programmed
frequency. In the calibration mode this indicator is not
illuminated for calibrating the output voltage. When it is
illuminated in the calibration mode the voltage measurement
can be calibrated.

I RMS

Illuminates when the LCD displays shows the programmed
current limit, measured current values, or current
measurement calibration.

I PK

Illuminates when the LCD displays shows the measured
peak current or peak current calibration. See Note.

PWR

Illuminates when the LCD displays shows the measured
power or power calibration. See Note.