Page 57

background image

User and Programming Manual - Rev R

2001RP

47

3. The output voltage must be programmed to 10.0 volts.

4. Use the SELECT key to select the Current Measurement calibration mode. This mode is

indicated by the I RMS LED.

5. Use the left control knob to adjust the current reading shown on the right hand LCD until

it reads as close as possible to 0.0. Do not go past a reading of zero.

1. Connect the test equipment to the power source as shown in

Low Scale Peak Current Measurement Calibration

Figure 6-1. Do not connect

the load.

2. Program the output voltage to 10.0 volts.

3. Use the SELECT key to select the Peak Current Measurement calibration mode. This

mode is indicated by the I PK LED.

4. Use the left control knob to adjust the peak current reading shown on the right hand LCD

until it reads as close as possible to 0.0. Do not go past a reading of zero.

1. Connect the test equipment to the power source as shown in

Low Scale Power Measurement Calibration

Figure 6-1. Do not connect

the load.

2. Program the output voltage to 10.0 volts.

3. Use the SELECT key to select the Power Measurement calibration mode. This mode is

indicated by the PWR LED.

4. Use the left control knob to adjust the power measurement reading shown on the right

hand LCD until it reads as close as possible to 0.0. Do not go past a reading of zero.

Note:

The method of adjusting the low scale values for voltage, current and
power insure that the measured value is correct for small values.
Under a no-load condition, however, the measured value may not be 0