AMETEK 2001RP User Manual
Page 57

User and Programming Manual - Rev R
2001RP
47
3. The output voltage must be programmed to 10.0 volts.
4. Use the SELECT key to select the Current Measurement calibration mode. This mode is
indicated by the I RMS LED.
5. Use the left control knob to adjust the current reading shown on the right hand LCD until
it reads as close as possible to 0.0. Do not go past a reading of zero.
1. Connect the test equipment to the power source as shown in
Low Scale Peak Current Measurement Calibration
Figure 6-1. Do not connect
the load.
2. Program the output voltage to 10.0 volts.
3. Use the SELECT key to select the Peak Current Measurement calibration mode. This
mode is indicated by the I PK LED.
4. Use the left control knob to adjust the peak current reading shown on the right hand LCD
until it reads as close as possible to 0.0. Do not go past a reading of zero.
1. Connect the test equipment to the power source as shown in
Low Scale Power Measurement Calibration
Figure 6-1. Do not connect
the load.
2. Program the output voltage to 10.0 volts.
3. Use the SELECT key to select the Power Measurement calibration mode. This mode is
indicated by the PWR LED.
4. Use the left control knob to adjust the power measurement reading shown on the right
hand LCD until it reads as close as possible to 0.0. Do not go past a reading of zero.
Note:
The method of adjusting the low scale values for voltage, current and
power insure that the measured value is correct for small values.
Under a no-load condition, however, the measured value may not be 0