4.3 Setting the Power on Initialization Values

The 2001RP is supplied with default factory settings when the unit is powered up. The factory
settings are:

Parameter

Factory default setting

Voltage range

Low

Auto Voltage Range

OFF

Voltage

5.0 Volt

Frequency

60 Hz ( 400 Hz on 2001RP-AV models)

Current limit

13.4 amps (17.4 amps on 2001RP-AV models, 14.8
amps on 2001RP-LZ models).

Display mode

Frequency

Output Relay state

OFF

Output start state

OFF

Output start phase

0 degrees

Local / Remote State

Local. Front panel unlocked.

Table 4-1: Factory Default Power on Settings

It is possible to change the power on initialization values in one of two ways:

1. Using the optional RS232 or IEEE-488 interface and the supplied PGUI32 program.

2. Using the front panel.

To change the power on initialization values from the front panel, proceed as follows:

1. Set the unit up in the desired way from the front. (Range, voltage, frequency, current

limit, output relay state, display mode).

2. Press and hold the SELECT key (normally toggles LCD display mode). Make sure the

correct indicator LED is illuminated, e.g., FREQ to display voltage and frequency.

3. While holding the SELECT key, press the OUTPUT ON/OFF key. This will save the

present front panel settings in non volatile memory register (NVM) no 7 and assign this
register as the power on register.

4. Release both keys. Note that as step 2 is executed, the display mode will toggle . This

means the display mode after power up is different from how the unit was set up. To
avoid this, just toggle the display till one position before the desired power on mode first
and then execute steps 2-4.

5. This procedure can be repeated as often as needed by the user.