10.5 Output Subsystem

This subsystem controls the main output function of the RP Series source.

Subsystem Syntax

OUTPut

[:RELay]

Opens or closes the output relay.

DROP

Drops the output voltage for a specified amount of time or
indefinitely.

STARt:

[STATe]

Enables or disables the start phase angle mode

PHASe

Sets the start phase angle value in degrees.

OUTPut[:RELay]

The output of the AC source can be isolated from the unit under test by opening the output
relay. The programmed output voltage is set to zero before opening the relay and is returned
to its programmed value when closing the relay. This avoids hot switching the output relay
contacts.

Command Syntax

OUTPut[:RELay]

Parameters

0 | 1

*RST Value

defined by the PONS

Examples

OUTP 1

Query Syntax

OUTPut?

Returned Parameters

0 | 1

Related Commands

OUTP:DROP

OUTP:STAR