2 current limit adjustment – AMETEK 2001RP User Manual

Page 60

background image

User and Programming Manual - Rev R

50

2001RP

4. Observe the actual output current as shown on the external DMM and current shunt.

Adjust R3 on the Range/Relay Board for a measurement display equal to the actual
output current.

CAUTION:

The amplifier wind tunnel must be covered for proper cooling if the
adjustment takes more than 1 minute.

6.4.2 Current Limit Adjustment

Note:

If the Current Limit and High Current Measurement Range Adjustments
are both adjusted, the High Current Measurement Range must be
adjusted first.

1. Connect the test equipment as shown in Figure 6-2. Use the 11.3 ohm load.

2. Program the current to 10.0 A. Program the output to 150V (135V for the -LZ option, 115

V for the –AV option) on the low voltage range and 60 Hz. (400 Hz for the –AV option).

CAUTION:

The amplifier wind tunnel must be covered for proper cooling if the
adjustment takes more than 1 minute.

1. Adjust R50 on the Oscillator Board for an output current of 10.5 A.