5 mkft (e3.1) technical data – BINDER MK 115 User Manual
Page 98
MKF + MKFT (E3.1) 07/2014
page 98/114
20.5 MKFT (E3.1) technical data
Unit size
115
240
720
Exterior dimensions
Width (including 18 mm / 0.7 in for 1 access port
(MKFT 115, 240), 36 mm / 1.4 in for 2 access
ports (MKFT 720), with plug)
mm / inch 1000 / 39.37 1135 / 44.69 1615 / 63.58
Height (incl. castors)
mm / inch 1725 / 67.91 1940 / 76.38 2005 / 78.94
Depth (incl. cable and door handle)
mm / inch 915 / 36.02 1000/ 39.37 1230 / 48.43
Depth (incl. cable and door handle) with optional
compressed air dryer
mm / inch 1085 / 42.72 1170 / 46.06 1400 / 55.12
Depth (incl. cable and door handle) with voltage
and frequency changer
mm / inch 1530 / 60.24 1615 / 36.58 1845 / 72.64
Wall clearance rear
mm / inch 300 / 11.81 300 / 11.81 300 / 11.81
Wall clearance rear with optional compressed air
dryer or to set up the voltage and frequency
changer
mm / inch 1000 / 39.37 1000 / 39.37 1000 / 39.37
Wall clearance sides
mm / inch
200 / 7.87 200 / 7.87 200 / 7.87
Window width
mm / inch 288 / 11.34 508 / 19.99 508 / 19.99
Window height
mm / inch
222 / 8.74 300 / 11.81 300 / 11.81
Number of doors
pc.
1
1
1
Interior dimensions
Width
mm / inch 600 / 23.62 735 / 28.94 1200 / 47.24
Height
mm / inch 480 / 18.90 700 / 27.56 1020 / 40.16
Depth
mm / inch 400 / 15.75 443 / 17.44 600 / 23.62
Interior volume
l / cu.ft.
115 / 4.06 228 / 8.05 734 / 25.92
Number of racks, standard / max.
pc.
1/4
1/6
1/11
Load per rack
kg / lbs.
30 / 66
30 / 66
40 / 88
Permitted total load
kg / lbs.
60 / 132
70 / 155
160 / 353
Temperature data (without humidity)
Temperature range 1)
°C / °F
-70 to +180 / -94 to 356
Temperature fluctuation 2)
± K
0.1 to 0.5
0.1 to 0.5
0.1 to 0.5
Temperature uniformity (variation) 2)
± K
0.1 to 1.3
0.2 to 1.8
0.3 to 2.0
Average heating up time acc. to IEC 60068-3-5
3)
K/min.
5.5
5.0
4.8
Average cooling down time acc. to IEC 60068-3-5
K/min.
4.2
4.2
4.0
Max. heat compensation up to 25 °C / 77°F
W
1500
3000
5000
Climatic data (with humidity)
Temperature range
°C / °F
+10 to +95 / 50 to 203
Temperature range
with optional compressed air dryer
°C / °F
0 to +95 / 32 to 203
Temperature fluctuation 2)
± K
0.1 to 1.0
0.1 to 1.5
0.1 to 1.0
Humidity range
% r.H.
10 to 98
10 to 98
10 to 98
Humidity range with optional compressed air dryer
% r.H.
5 to 98
5 to 98
5 to 98
Humidity fluctuation 2)
± % r.H.
≤ 2.5
≤ 2.5
≤ 2.5
Dew point temperature range
°C / °F
+5 to +94 / 41 to 201
Dew point temperature range
with optional compressed air dryer
°C / °F
-28 to +94 / -18.4 to 201
Max. heat compensation up to 25 °C / 77°F at 90
% r.H. (without optional compressed air dryer)
W
400
400
800