10. Bedew protection facility (Operation line 1)

When operating the unit without humidification, the bedew protection condensates the chamber humidity

at the coldest point in order to avoid the samples becoming wet by condensation. Bedew protection is

performed by the evaporator and can be programmed On/Off via operation line 1 in Manual Mode

(HAND) and in Program Mode (AUTO).

Use the bedew protection only if it is necessary to prevent condensation at the charging mate-

rial.

Use the bedew protection only when operating the unit without humidification.

When the bedew protection is enabled (operation line 1 = On) the refrigeration machine keeps operating

within warming-up phases (On = refrigeration machine operating, Off = refrigeration machine off).

If possible, use the bedew protection only during warm-up phases. If necessary, you can also

activate it during hold phases.

Do NOT use the bedew protection above a temperature set-point of +20 °C / 68°F maximum.

To obtain optimal warming results without condensation on the samples, program a heating gradient of

approx. 0.5 °C/min.

Example:

W/°C

t/min.

00

01

02

03

-40

-20

0

20

40

60

80

120

180

380

04

ON

OFF

t/min.

Minimum

tolerance

Maximum

tolerance

Operation line 1 = bedew protection


In the program table, the state of opera-

tion line 1 for bedew protection is repre-

sented in the column “Sk”:
On = 0000000

1

Off = 0000000

0

Example:
Operation line 1 is activated in the first

program section and remains active up

to 20 °C / 68°F.
Then the refrigeration machine turns off

automatically.

PGM

08:43:55 15.12.13