7 operating instructions, 8 measures to prevent accidents – BINDER MK 115 User Manual
Page 21

MKF + MKFT (E3.1) 07/2014
page 21/114
The charging material shall not contain any corrosive ingredients that may damage the ma-
chine components made of stainless steel, aluminum, and copper. Such ingredients include in
particular acids and halides. Any corrosive damage caused by such ingredients is excluded
from liability by BINDER GmbH.
1.7 Operating instructions
Depending on the application and location of the unit, the operator of the alternating climate chambermust
provide the relevant information for safe operation of the unit in a set of operating instructions.
Keep these operating instructions with the unit at all times in a place where they are clearly
visible. They must be comprehensible and written in the language of the employees.
1.8 Measures to prevent accidents
The operator of the alternating climate chamber MKF / MKFT must observe the following rule: “Betreiben
von Arbeitsmitteln. Betreiben von Kälteanlagen, Wärmepumpen und Kühleinrichtungen“ (Operation of
work equipment. Operation of refrigeration systems, heat pumps and refrigeration equipment) (GUV-R
500 chap. 2.35) (for Germany).
The manufacturer took the following measures to prevent ignition and explosions:
•
Indications on the type plate
See operating manual chap. 1.4.
•
Operating manual
An operating manual is available for each alternating climate chamber.
•
Overtemperature monitoring
The alternating climate chamber is equipped with a temperature display, which can be read from outs.
The unit is equipped with an additional safety controller (temperature safety device class 2 acc. to DIN
12880:2007). Visual and audible (buzzer) signals indicate temperature exceeding.
•
Safety, measurement, and control equipment
The safety, measuring, and control equipment is easily accessible.
•
Electrostatic charge
The interior parts are grounded.
•
Non-ionizing radiation
Non-ionizing radiation is not intentionally produced, but released only for technical reasons by electri-
cal equipment (e.g. electric motors, power cables, solenoids). The machine has no permanent mag-
nets. If persons with active implants (e.g. pacemakers, defibrillators) keep a safe distance (distance of
field source to implant) of 30 cm, an influence of these implants can be excluded with high probability.
•
Protection against touchable surfaces
Tested according to EN ISO 13732-1:2008.
•
Floors
See operating manual chap. 3.4 for correct installation
•
Cleaning
See operating manual chap. 17.2.