Page 28
MKF + MKFT (E3.1) 07/2014
page 28/114
3.2 Guidelines for safe lifting and transportation
The front castors of the alternating climate chamber can be blocked by brakes. Please move the units
with castors only when empty and on an even surface, otherwise the castors may be damaged. Mount
the upholstered steel L-type profile at the lower door locking. After operation please observe the guide-
lines for temporarily decommissioning the unit (chap. 18.2).
CAUTION
Sliding or tilting of the unit.
Damage to the unit.
Risk of injury by lifting heavy loads.
Do NOT lift or transport the unit using the door handle, the door or at the lower hous-
ing.
Do NOT lift the unit by hand.
Transport the unit only in its original packaging.
Secure the unit with transport straps for transport.
Keep the unit in upright position.
Place the unit using technical devices (fork lifter) on the transport pallet. Set the fork
lifter only from the rear in the middle of the unit. Make sure to place all the lateral sup-
ports of the unit on the forks.
Transport the unit with the original transport pallet. Set the fork lifter ONLY to the pal-
let. Without the pallet the unit is in imminent danger of overturning.
You can order transport packing and pallets for transportation purposes from BINDER service.
Permissible ambient temperature range during transport:
•
If the steam humidifying system has NOT been emptied: +3 °C / 37.4 °F to +60 °C / 140 °F.
•
After BINDER Service has emptied the steam humidifying system: -10 °C / 14 °F to +60 °C / 140 °F.
With temperatures below +3 °C / 37.4°F, water must be completely removed from the humidifying system.
CAUTION
Transport below +3 °C / 37.4°F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the unit.
Contact BINDER Service before any transportation below +3 °C / 37.4°F.
3.3 Storage
Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for tempo-
rary decommissioning (chap. 18.2).
Permissible ambient temperature range during storage:
•
If the steam humidifying system has NOT been emptied: +3 °C / 37.4 °F to +60 °C / 140 °F.
•
After BINDER Service has emptied the steam humidifying system: -10 °C / 14 °F to +60 °C / 140 °F.