Soapsuds may contain chlorides and must therefore NOT be used for cleaning.

With every cleaning method, always use adequate personal safety controls.

Following cleaning, leave the unit door open or remove the access port plugs.

The neutral cleaning agent may cause health problems in contact with skin and if ingested.

Follow the operating instructions and safety hints labeled on the bottle of the neutral clean-

ing agent.

Recommended precautions: To protect the eyes use sealed protective goggles. Suitable protective

gloves with full contact: butyl or nitrile rubber, penetration time >480 min.

CAUTION

Contact with skin, ingestion.
Skin and eye damage due to chemical burns.

Do not ingest. Keep away from food and beverages.

Do NOT empty into drains.

Wear protective gloves and goggles.

Avoid skin contact.


17.2.2 Decontamination

The operator must ensure that proper decontamination is performed in case a contamination of the

chamber by hazardous substances has occurred.
Disconnect the chamber from the power supply prior to chemical decontamination. Pull the power plug.
Do not use decontamination agents that may cause a hazard due to reaction with components of the

device or the charging material. If there is doubt regarding the suitability of cleaning products, please

contact BINDER service.
You can use the following disinfectants:

Inner chamber

Standard commercial surface disinfectants free from acid or halides.
Alcohol based solutions.
We recommend using the disinfectant spray Art. No. 1002-0022.

For chemical disinfection, we recommend the disinfectant spray Art. No. 1002-0022.
Any corrosive damage that may arise following use of other disinfectants is excluded from

liability by BINDER GmbH.

With every decontamination method, always use adequate personal safety controls.