4.2.1 Automatic fresh water supply for humidifying system via water pipe

An enclosure inside the unit contains the connection kit for water supply and wastewater. Install the water

supply connection using either the enclosed water hose or another pressure-resistant one. To accomplish

this, remove the cover of the freshwater connection “IN” (18) (Figure 8) on the rear of the unit. Protect

both ends of the hose with two of the four supplied hose clamps. Before turning on the unit, check the

connection for leaks. Water supply is automatically effected via the freshwater connection “IN” (18).

As the unit only lets in water when required, there is no continuous water flow.

Supply pressure 1 to 10 bar when connecting to a water pipe.

Water type: deionized (demineralized) water

Water intake temperature NOT below +5 °C / 41 °F and not exceeding 40 °C / 104 °F.

The water intake shall be provided with a shut-off slide or water-tap.

For the water supply, fix the delivered adapter with hose olive on the thread at the rear of

the chamber.

Protect the water supply at one side with the supplied hose clamp.


4.2.2 Manual fresh water supply for humidifying system via internal freshwater can

If no house water connection with suitable water is available, you can manually supply water by filling the

freshwater can (total volume: 19 liters / 0.67 cu.ft. up to the maximum level mark), which is located behind

the right door of the humidity generation module.
The cover of the water inlet valve must be screwed on the freshwater connection “IN” (18). Open the door

(F) (Figure 3) to access the filler neck of one of the water can. You cannot totally take out the water can

because of its fix connections. Fill the water can only up to ¾, up to the maximum level mark. When filling

it too much with the chamber turned on, the alarm message “

WATER LEVEL TOO HIGH

” is displayed on the

controller (chap. 13.1). Manually suck off the water, or operate the unit with high temperature and humidi-

ty values until the excess water is consumed. When filling it too much with the chamber turned off, water

can escape from the unit. Thus, ensure not to fill the can by more than the maximum level mark.

To guarantee humidification during 24 hours even at high humidity set-points with manual

water supply, we recommend filling the freshwater can (option) daily at the end of the day.