BINDER MK 115 User Manual
Page 72
MKF + MKFT (E3.1) 07/2014
page 72/114
Checking and setting safety controller set-point type and safety controller set-point:
Unlock the keyboard locking (option, chap. 16.5).
Grundeinstellung
User-settings
Instrument data
Contrast
Displ. Power down
Contin. operation
User Level
Safety control. Act + 100,8 °C
Safety control. Set. + 105.0 °C
29
User level code no.
1 (factory setting)
In the menu “User Level” select the submenu “Safety controller”.
•
Select the safety controller set-point type “Limit” or “Offset” in the field “Setpoint type”
•
Enter the value for “Limit” or “Offset” in the field “Setpoint”.
Lock afterwards the keyboard locking (option, chap. 16.5).
For temperature disturbances see alarm indications, chap. 13.
User Level
Date
and time
Summer time
Language
Temperature unit
Buzzer
Safety controller
User-code No.
English
°C
1
Active
Safety Controller
Setpoint type
Setpoint
+ 110.0
Limit
Sollwert Art
Grenzwert
Offset
Setpoint
+0110.0
Sollwert Art
Grenzwert
Offset
Setpoint type
Limit
Offset
4 x