Notification and alarm functions – BINDER MK 115 User Manual
Page 74
MKF + MKFT (E3.1) 07/2014
page 74/114
13. Notification and alarm functions
13.1 Notification and alarm system overview (auto diagnosis system)
•
Visual indications of notification or error messages are blue notes on the display of the MB1controller.
•
Visual indications of an alarm messages are red notes with an alarm bell symbol.
In addition, there is an audible alert, if you did not deactivate the buzzer in the “User level” menu
(chap. 6.4).
Event
Note
(blue field)
Alarm
(red field)
Lack of water in the freshwater can. The unit
continues working for a few hours.
In case of freshwater supply via water pipe,
the water tap is closed, or the unit is defective.
WATER LOW
after 5 min
Wastewater pump or float switch in the
wastewater can defective (contact BINDER
service), or wastewater hose clogged.
WASTEWATER
after 10 sec.
Humidifying unit defective. Contact BINDER
service.
FAULT STEAM SYSTEM
immediately
Freshwater can is too empty to allow normal
function, or the unit is defective.
The humidification system turns off.
Fill up the water can or open the water supply.
During the one-hour preheating phase: mes-
sage without significance
WATER TANK EMPTY
after 60 sec.
When manually filling up the freshwater can:
Freshwater can is too full, or the unit is defec-
tive. Suck off the water (chap. 4.2.2). The unit
functions as usual.
If the message persists, contact BINDER ser-
vice.
WATER LEVEL TOO
HIGH
after 60 sec.
Humidity switch OFF or humidity outside con-
trol range. No active humidification of the inte-
rior. Further dehumidification possible by re-
frigerating operation
HUMID OFF
immediately
One-hour preheating phase, no refrigerating
and dehumidification functions
1H PREHEAT PHASE
immediately
Fault in refrigerating machine. Contact
BINDER service.
FAULT COMPRESSOR
immediately