BINDER MK 115 User Manual
Page 55
MKF + MKFT (E3.1) 07/2014
page 55/114
Enter the individual values of the selected program section.
Program editor Abs.Nr. 6
****.*
Setpoint 1
+100.0
FAN
Operating contacts
Time
00:45:00
Repeat Section
5
Repeat Number
10
Tol.-band min.
-1999.0
Tol.-band max.
+9999.0
Parameter set
1
Setpoint 1
Temperature at the start of the program section
FAN
Fan speed in % (no function)
Operating contacts Operation lines ON/OFF
Time
Duration of the program section
Repeat Section
No. of start section in case of repeat cycles
Repeat Number
No. of repeats in case of repeat cycles
Tol.-band min.
Temperature limits (maximum / minimum temperature)
(In case of exceeding: temporary program stop)
Tol.-band max.
Parameter set
Pre-selected value (Do NOT change!)
Select the parameters via the arrow keys and confirm by pressing ENTER.
Then enter the values via the arrow keys, and confirm the entry by pressing ENTER.
For a negative set-point entry, enter the numerical value first, and then the minus sign (-).
With set-point type “
Limit”, the user shall adapt the safety controller (chap. 12.2) to the high-
est temperature set-point value of the program actually used. Check the safety controller for
each temperature program and change it if necessary. Set the safety controller set-point by
approx 10 °C above the highest temperature set-point of the program.
In case of the optional over-/under temperature safety device (chap. 12.3), check also the
temperature limits entered there, and adjust them if necessary.
Performance after completing the program:
The controller changes to Idle Mode. Heating, refrigeration, and humidification are inactive; the chamber
approximates ambient temperature.
The fan is off. The switching states of the operation lines are OFF.