7.
Graphic representation of the historical measurement (chart
recorder function)
The representation of data imitates a chart recorder and allows recalling any set of measured data of any
point of time taken from the recorded period.
Normal display of the chart recorder function:
Top left: The actual date and time are displayed.
Below: The current values of temperature [ °C] and humid-
ity [% r.H.] are numerically and graphically displayed.
Scaling:
MKF: Temperature: -50 °C / -58°F to +200 °C / 392°F
MKFT: Temperature: -100 °C / -148°F to +200 °C / 392°F.
Humidity: 0% r.H. to 100% r.H.
Button
permits toggling between different repre-
sentations.
Depending on the selected kind of representation, button
H
might not have been visible until this procedure.
Activation of the optional over- or under temperature safety device (chap. 12.3) is displayed on the right
side of the display as an enlarged blue line.
The active bedew protection is displayed on the right side of the display as an enlarged green line.
History display with cursor:
Select button
H
= History. A pink line appears on the
display marking as a cursor the selected moment. You
can now recall the recorded data of any defined moment.
Top left: Date and time of the selected cursor position are
displayed.
Below: The corresponding temperature and humidity val-
ues of this instance are numerically and graphically dis-
played.
Scroll the cursor position using the arrow buttons.
Single arrow buttons: fine-tuning.
Double arrow buttons: page-up and page-down.
Switch to the zoom display by pressing button
:
History - zoom function:
Magnifier buttons
: Zoom and zoom
back (i.e., shorten or extend the displayed period).
Toggle back to the former representation display using
this button
.
11:32:37 15.12.13
+22.6
% rH
11:32:20 15.12.13
10:45:00
+23.9 °C
+22.6
% rH
11:05:00
11:32:14 15.12.13
10:45:00
+23.9 °C
CONFIG
VIEW->
H
+22.6 % rH
11:05:00