2 function overview program controller mb1, Ole_link28 – BINDER MK 115 User Manual
Page 45
MKF + MKFT (E3.1) 07/2014
page 45/114
6.2 Function overview program controller MB1
The main operation level contains the following dif-
ferent displays:
•
Normal display (Idle Mode or Manual Mode or
Program Mode)
•
Event List
•
Chart recorder function
•
Contact page
Toggle between the displays by pressing button
VIEW ->
.
The
NORMAL DISPLAY enables comparison of the
current temperature and humidity (W) to the set-
point values (X).
CONTACT PAGE
BINDER Service contact display.
NORMAL DISPLAY Idle Mode
W
X
08:43:55 15.12.13
CONFIG
VIEW->
PGM
TEMP
0.0
26.8
°C
HUMID
0.0
58.6
%rH
HAND
or
W
X
08:43:55 15.12.13
CONFIG
VIEW->
PGM
X - TEMP
26.8
°C
X - HUMID
58.6
%rH
HAND
No heating or refrigeration, no humidification or dehumidification. The actual values (X) approximate ambi-
ent temperature and humidity. The fan is off.
NORMAL DISPLAY Manual Mode
W
X
08:43:55 15.12.13
CONFIG
VIEW->
PGM
TEMP
40.0
36.8
°C
HUMID
65.0
58.6
%rH
HAND
Temperature and humidity values are maintained
according to the previous entered set-points (W).
NORMAL DISPLAY Program Mode
W
X
08:43:55 15.12.13
PROGRAM 01/SEC1 00:09:59
CONFIG
HAND VIEW->
PGM
TEMP
40.0
36.8
°C
HUMID
65.0
58.6
%rH
AUTO
A temperature and humidity program entered before
via a program table is run.
EVENT LIST
Overview over the last 16 events or error occur-
rences of the unit.
CHART RECORDER FUNCTION
Graphical display of the current temperature and hu-
midity values and review of the previous measure-
ments on a historical display. A memory interval of 5s
corresponds to a supervision period of 2.5 days.
11:32:14 15.12.13
10:45:00
+23.9 °C
CONFIG
VIEW->
H
+22.6 % rH
11:05:00
08:43:55 15.12.13
08:43:55 15.12.13
Service Hotline
International: +49 7462 205 555
USA Toll Free: +1 866 885 9794
or +1 631 224 4340
: +7 495 988 1517
РоссияиСНГ
www.binder-world.com
CONFIG
VIEW->
VIEW->
Best conditions for your success