beautypg.com

8 electrical connection, Caution – BINDER MK 115 User Manual

Page 39

background image

MKF + MKFT (E3.1) 07/2014

page 39/114


For the installation of the voltage and frequency changer behind the alternating climate chamber, provide

a rear wall distance the alternating climate chamber of approx. 1 m.
If possible, fix the voltage and frequency changer at the alternating climate chamber. For this purpose, an

Allen key size 4 is required. Connect the slots at the end of the chassis with two M6 screws to the threads

provided below on the rear panel of the alternating climate chamber (see Figure 13).

CAUTION

Danger of overheating.
Damage to the voltage and frequency changer.

Do NOT install the voltage and frequency changer in unventilated recesses.

Ensure sufficient ventilation for dispersal of the heat.

The voltage and frequency changer is equipped with

four castors. The rear castors can be easily locked

via the attached brakes

4.8 Electrical connection

4.8.1 Information on connecting the alternating climate chamber

The alternating climate chamber MKF / MKFT comes with a fixed power connection cable 2700 mm / 8.9

ft in length and 3 internal overload releases against excess-current. The socket must also provide a pro-

tective conductor.

Power plug

Voltage

+5% /-10 %

Current

type

Power

frequency

Unit fuse

MKF 115

MKFT 115

MKF 240

CEE plug 5-poles, 16 Amp

400 V

3 N~

50 Hz

16 Amp

3 x internal

MKFT 240 CEE plug 5-poles, 16 Amp

400 V

3 N~

50 Hz

16 Amp

3 x internal

MKF 720

MKFT 720 CEE plug 5-poles, 32 Amp

400 V

3 N~

50 Hz

25 Amp

3 x internal

Prior to connection and start-up, check the power supply voltage. Compare the values to the specified

data located on the unit’s type plate (left unit side, bottom right-hand, chap. 1.4)
When connecting, please observe the regulations specified by the local electricity supply company and as

well as the VDE directives (for Germany). We recommend the use of a residual current circuit breaker.

Pollution degree (acc. to IEC 61010-1): 2

Over-voltage category (acc. to IEC 61010-1): II

CAUTION

Danger of incorrect power supply voltage.
Damage to the equipment.

Check the power supply voltage before connection and start-up.

Compare the power supply voltage with the data indicated on the type plate.

See also electrical data (chap. 20.4).

To completely separate the unit from the power supply, you must disconnect the power plug.

Install the unit in a way that the power plug is easily accessible and can be easily pulled in