2 resetting the notification or alarm messages, Caution – BINDER MK 115 User Manual
Page 75
MKF + MKFT (E3.1) 07/2014
page 75/114
Event
Note
(blue field)
Alarm
(red field)
Operation line 1 (bedew protection) activated
DRY
immediately
Limit value of safety device exceeded
TEMPERATURE LIMIT
immediately
With option over/under temperature safety device class 2 (chap. 12.3):
Exceeding the maximum / minimum tempera-
ture
TEMP LIMIT
immediately
With option keyboard locking (chap. 16.5):
Locked keyboard
KEY LOCK
immediately
The indicated intervals refer to the time after occurrence of the error or notified condition.
13.2 Resetting the notification or alarm messages
The “RESET” button, which serves to acknowledge and reset the indication, will become visible automati-
cally whenever a notification or an alarm message appears.
1. Depending on the type of error, eliminate the cause of the alarm or wait until the unit compensates for
the reason of the error.
2. Press the “RESET” button to reset the notification or alarm message.
CAUTION
In case the “RESET” button does not cancel the notification or alarm indication, the reason
for the disturbance was not removed correctly.
Contact BINDER service.