Ole_link2, Ole_link3, Ole_link4 – BINDER MK 115 User Manual
Page 97: Ole_link5
MKF + MKFT (E3.1) 07/2014
page 97/114
Unit size
115
240
720
Electrical data
IP-system of protection acc. to EN 60529
IP
20
20
20
Nominal voltage (+/-10%) 50 Hz
V
400 3N~
400 3N~
400 3N~
Nominal power
kW
4.50
6.80
11.00
Energy consumption
at +25 °C / 77 °F and 60 % r.H.
Wh/h
1250
1500
3900
Power plug: CEE plug 5-poles
Amp
16
16
32
Over-voltage category acc.
to
IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Over-current release category B,
3 x internal
Amp
16
16
25
Electrical data of the voltage and frequency changer
IP-system of protection acc. to EN 60529
IP
23
23
23
Nominal voltage (+/-10%) 60 Hz (input side)
V
480 3N~
480 3N~
480 3N~
Nominal power
kW
9
9
13
Over-voltage category acc.
to
IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Fuse
A
16
16
25
Further information
Weight (empty)
kg / lbs.
280 / 617
360 / 794 590 / 1300
Weight (empty) with optional compressed air dryer
kg / lbs.
295 / 650
375 / 827 605 / 1334
Filling weight of refrigerant R 404A (GWP 3750)
kg / lbs.
2.00 / 4.41 2.20 / 4.85 5.00 / 11.02
Noise level
approx. dB(A)
62
65
65
Noise level with optional compressed air dryer
(short-term)
approx. dB(A)
85
85
85
Noise level with voltage and frequency changer
approx. dB(A)
67
67
67
1)
Lower values are valid at an ambient temperature of max. 25 °C / 77°F
2)
Depending on the set-point
3)
Units with voltage and frequency changer: Values reduced by 0.3 K/min each.
*
Data not yet determined
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
+25 °C / 77°F and a power supply voltage fluctuation of +/-10%. The temperature data is determined in
accordance to BINDER factory standard following DIN 12880, observing the recommended wall clear-
ances of 10 % of the height, width and depth of the inner chamber.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.
Bringing a source of humidity into the inner chamber will affect the minimum humidity specifi-
cation and may affect the humidity accuracy.