beautypg.com

Manual mode, 1 set-point entry, Ole_link1 – BINDER MK 115 User Manual

Page 51

background image

MKF + MKFT (E3.1) 07/2014

page 51/114

8.

Manual Mode

In Manual Mode (HAND) you can enter a temperature set-point, a humidity set-point, and the switching-

state of up to 8 operation lines. Operation line 1 can be used to control the bedew protection (chap. 10).

Operation lines 2 to 5 serve to switch any device connected to the zero-voltage relay outputs (DIN sock-

ets (11) and (12), chap. 11). Operation line 8 releases the compressed air dryer (option, chap. 16.7). The

other operation lines are non-functional. All settings remain valid in Manual Mode (HAND) until the next

manual change, if the unit had been turned off or in case of toggling to Idle Mode or Program Mode

(AUTO).

8.1 Set-point entry


Idle Mode
No heating or refrigerating function,

no humidification.
The fan is off.

Toggling to Manual Mode


Change of page

Toggling between temperature set-point, humidity set-point, and operation lines.

Arrow buttons to select the operation

lines

(Operation line 1 = bedew protection)

Arrow buttons to

enter the value

Button to move the decimal point

Unlock the keyboard locking (option, chap. 16.5) via the key switch to enter the set-point.

08:43:55 15.12.13

HAND

08:43:55 15.12.13

Operating contacts

DRY

Contact

2

Contact

3

Contact

4

Contact

5

Contact

6

Contact

7

Contact

8

On

Off

Aus

Aus

Aus

Aus

Aus

Off

HAND

DRY

Off

On

Hand-Mode

TEMP

Oper

+

40

.0 °C

+80.0 %

TEMP

+0060.0

°C

HUMID

+65.0 %r.H.

EXIT

08:43:55 15.03.13

HAND

HAND

Hand-Mode

TEMP

Steuerkontakte

+29.3 °C

+80.0 %

Hand-Mode

TEMP

Operating contacts

+40.0 °C

PGM

CONFIG

W

VIEW->

EXIT

HUMID

+65.0 %r.H.

W

X

08:43:55 15.12.13

CONFIG

VIEW->

PGM

TEMP

0.0

26.8

°C

HUMID

0.0

58.6

%rH

HAND