8 performance after power failure in program mode – BINDER MK 115 User Manual
Page 63
MKF + MKFT (E3.1) 07/2014
page 63/114
9.7 Repetition of a section or several sections within a program
Here we use the example of a set-point ramp temperature program of chap. 9.5. The shaded sections 02
and 03 shall be repeated e.g. 30 times.
Program
section
No.
Set-
point
temp.
W-1
Fan
FAN
Section time
Time
Operation lines
Sk
Target
section
No
No. of
cycles
Cy
Minimum
tolerance
Tmin
Maximum
tolerance
Tmax
01
-20
****.
00:30:00
00000000
1
0
-1999
+9999
02
0
****.
01:30:00
00000000
1
0
-5
+5
03
60
****.
01:00:00
00000000
1
0
-2
+2
04
60
****.
03:20:00
00000000
1
0
-1999
+9999
05
-40
****.
00:00:01
00000000
1
0
-1999
+9999
The following table shows the program that results, whereby the differences to the table above are shad-
ed.
Program
section
No.
Set-
point
temp.
W-1
Fan
FAN
Section time
Time
Operation lines
Sk
Target
section
No
No. of
cycles
Cy
Minimum
tolerance
Tmin
Maximum
tolerance
Tmax
01
-20
****.
00:30:00
00000000
1
0
-1999
+9999
02
0
****.
01:30:00
00000000
1
0
-5
+5
03
60
****.
01:00:00
00000000
2
30
-2
+2
04
60
****.
03:20:00
00000000
1
0
-1999
+9999
05
-40
****.
00:00:01
00000000
1
0
-1999
+9999
Sections 02 and 03 will be executed in total 31 times; only then will the program continue.
Entry of the values into the display program table:
To have sections repeated indefinitely, enter the number of cycles “Cy” as -1.
9.8 Performance after power failure in Program Mode
The program is resumed at the point where the interruption occurred with the latest set-points reached
during the program run. The power failure is noted in the event list. No error message is displayed indicat-
ing that a power failure had taken place.