AIN 1 = Plus from CMPupw
AIN 2 = Plus from CMPdown
AIN 3 = Plus from CGRupw
AIN 4 = Plus from CGRdown
AIN 6 = common ground for all above instruments.

Vref = 10K thermistor
AIN 5 = 10K thermistor & 10K reference resistor
GND = 10K reference resistor.

The next illustration displays the software configuration:


Please do not forget to enter the Polynomials P2 - P5 inside the coefficient area.

And please be aware of the multiplexing principle inside the Logbox SD, it will take at least a full
second before each measurement is performed. Meaning there will be a 4second time shift
between the measurement taken of the first channel and the measurement taken on the fourth
channel. (Albedo measurements should be taken on the same time interval.)