beautypg.com

Kipp&Zonen LOGBOX SD Data Logger User Manual

Page 31

background image




31

12 POLYNOMIALS CALCULATION FOR SENSORS


This section will explain how to change the polynomial to be applied when you are using radiometers
inside the Logbox SD graphical user interface.


Each pre-defined radiometer has a default Polynomial. (A1) This could be set to100 or 10, and needs to be
changed to have a correct reliable output in W/m².

Step1 : Look at your radiometer, and locate the sensitivity. This is located on the radiometer itself and on
the calibration certificate. (Example: 22,14µV/W/m

2

)


Step2: Calculate your own polynomial by applying the next formula : A1=1000/sensitivity
The reason for his is that the Logbox SD is measuring in mV and not in µV.

Step 3: Change your polynomial with the outcome of the formula. (1000/22,14µV/W/m

2

= 45.16)