Safety Information

20

FG-750

Electrical Safety Information

W

ARNING

Use this unit indoors only.

Position the unit so that air can circulate freely around it.

Operation of any electrical instrument around flammable gases
or fumes constitutes a major safety hazard.

To avoid electrical shock, do not operate the unit if any part of
the outer surface (covers, panels, etc.) is damaged.

Capacitors inside the unit may be charged even if the unit has
been disconnected from its electrical supply.

DO NOT connect the unit interfaces metallically to OSP (Outside
Plant) wiring. The unit interfaces are designed for use as intra
building surfaces only (Type 2 or Type 4 ports as described in
GR- 1089-CORE) and require isolation from the exposed OSP
cabling. The addition of Primary Protectors is not sufficient
protection in order to connect these interfaces metallically to
IOSP wiring.