EXFO FG-750 User Manual

Page 126

background image

Setting Up Your RTU

112

FG-750

Detecting the Fibers Connected to the Optical Ports

3. From the displayed dialog box, select

a wavelength.

Note: The wavelength used to detect a series of

ports will be used as the default
monitoring test wavelength. You may
therefore start monitoring on certain
ports at a certain wavelength and on
other ports with another wavelength.

4. Select the ports you want to use at

this wavelength.

Note: To quickly select or clear the check boxes, use respectively Select All and

Unselect All.

5. Click Start to start the automatic detection of the ports. The RTU will

generate test setups for the new wavelengths.

6. Repeat the previous steps with other wavelengths and ports if

necessary.

The status of the ports (detected or not) is indicated in the OTAU
section.

You can now configure the remote switch, if you have one
(see Configuring a Remote Switch (ROTAU) on page 115).

Note: Once the manual detection of fiber ports is done there is a chance that ports

remain undetected though the fiber is connected, these ports can be
detected forcefully by selecting force detection of fiber.