6 animal weighing function – KERN & SOHN ALT 250-4B User Manual
Page 84

84
ALT_B-PLT_A-PLT_F-BA-e-1333
12.6 Animal weighing function
The animal weighing function can be applied for busy weighings. During a defined
period the mean value of the weighing results is formed.
The more unquiet the weighed item, the longer the measuring time should be
selected.
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons
to select the menu item „Animal weigh.“.
Confirm with the PRINT button, the currently set measuring time will be displayed.
Use the navigation buttons to select the wanted measuring time (5
– 90 s) and
confirm with PRINT button.
Put upon weighing goods and press the PRINT button. In the display runs a
„Countdown“ of the set measuring time.
The average value of the weighing results is displayed and remains displayed on
the screen.
When an optional printer is connected, the displayed value can be edited using
PRINT. Printout example (KERN YKB-01N):
07-09-11 11:14:57
Time
= 20 sec
Measuring time
A:
20.0052 g
Weighing result
For other weighing procedures press ON/OFF button 1x
Return to weighing mode:
Press the ON/OFF button 2x