Setting adjustment mode:
In weighing mode press the MENU button and keep it pressed until the
acoustic signal gets mute. The setup menu is displayed.
Use the navigation buttons
to select the menu item „Adjustment mode“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
Display
Description
Procedure see chap.
Internal adjustm.
Adjustment with internal
adjustment weight
Chap. 7.1
External adjustm.
Adjustment with external
adjustment weight
Chap. 7.3
Technical adjustm.
Overwrite internal
adjustment weight
Chap. 7.4
Autom. adjustm.
Automatic adjustment
with internal adjustment
weight
Chap. 7.2
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.