12.5 Percent determination
Percent determination allows weight display in percent, in relation to a reference
weight.
12.5.1 Entering the reference weight by weighing
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu i
tem „Percent weigh.“.
Confirm with the PRINT button, the currently set mode will be displayed.
Use the navigation buttons
to select setting „Automatic“.
If using a weighing container, put this before pressing the PRINT button. The
taring process is automatically started.
Press the PRINT key.
Place the reference weight (=100 %) and press the PRINT button. As soon the
weighing value is stable, the determined weight will be taken over as reference
(100%). From now the balance is in percent determining mode.
Place weighed goods, the result display will appear.
W%
Weight of the sample in percentage
W
Weight of the sample in grams
REF: Reference weight (100%)
When an optional printer is connected, the displayed value can be edited
using PRINT. Printout example (KERN YKB-01N):
07-09-11 11:14:57
Perc.
49.95 %
Weight of the sample in
percentage
Weight:
9.990 g
Weight of the sample in grams
Refer.:
20.000 g
Reference weight (100%)