P.12.4 – KERN & SOHN ALT 250-4B User Manual
Page 79

ALT_B-PLT_A-PLT_F-BA-e-1333
79
12.4 Check weighing
In checkweighing an upper and a lower limit value can be established to ensure that
the weighed goods are exactly within the established tolerance limits.
The tolerance mark () and an acoustic signal (selectable) show, if the weighed
material is within the two tolerance limits.
Display of the results
1. Lower and upper limit value
0
Tolerance
marker
Tone
Assessment
no
Load exceeds specified tolerance
yes
Load within specified tolerance
no
Load below specified tolerance
2. Lower limit value > 0 and upper limit value = 0
Tolerance
marker
Tone
Assessment
yes
Weighed goods > lower limit value
no
Weighed goods
lower limit value
3. Lower limit value = 0 and upper limit value > 0
Tolerance
marker
Tone
Assessment
no
Weighed goods > upper limit value
yes
Weighed goods
upper limit value