8.1 Automatic adjustment with internal weight
The automatic adjustment with internal weight
is started
automatically, if the balance
after the weighing balance was disconnected from the mains
after pressing ON/OFF in stand-by mode
after a temperature change of 1.5 °C
with non loaded weighing pan / zero display
after a time interval of 20 min with non loaded weighing pan / zero
display
The automatic adjustment function is always enabled. At any time, however, you can
carry out a manual adjustment (press CAL-button) using the internal adjustment
weight, see chap. 7.1.
Or selecting the adjustment mode in the menu:
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons
to select the menu item „Adjustment mode“.
Confirm by the PRINT button
, „Autom. Just.“ will be displayed.
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.