ALT_B-PLT_A-PLT_F-BA-e-1333
7.3 Adjustment with external adjustment weight
Condition:
Menu setting „External adjustm.“, see chap. 7
At the models with internal adjustment weight the adjustment with
external weight is not recommended.
Information about test weights you will find in the internet under
Ensure that there are no objects on the weighing pan. In weighing mode press
the CAL button.
Wait until the weighed value for the required adjustment weight appears
flashing.
During the flashing display put the required adjustment weight carefully in the
center of the weighing pan. The flashing display disappears.
After successful adjustment the balance automatically returns to weighing
mode.
Take away adjustment weight