11.4 Auto Zero
Under this menu item the automatic zero point correction can be switched on or off.
In switched-on-state the zero point is automatically corrected at drift or when dirty.
Note:
In the event that small quantities are removed or added to the material to be
weighed, incorrect weighing results can be displayed due to the “stability
compensation“. (e.g. slow flow of liquids from a container placed on the balance,
evaporating processes).
When apportioning involves small variations of weight, it is advisable to switch off
this function.
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons
to select the menu item „Auto Zero“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
! In models PLT-M
only „Auto Zero Off“ and „Auto Zero 1“ are available.
Display
Description
Auto Zero off
Auto Zero switched off
Auto Zero 1
Auto Zero range ± ½ Digit
Auto Zero 2
Auto Zero range ± 3 Digits
Auto Zero 3
Auto Zero range ± 7 Digits
Auto Zero 3E
Auto Zero range ± 7 digits in the whole weighing range
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.