KERN & SOHN ALT 250-4B User Manual
Page 37

ALT_B-PLT_A-PLT_F-BA-e-1333
37
Contrast adjustment
(see chap. 11.7)
1-15
Setting the contrast of the display
Backlight (see chap. 11.8)
on
Background illumination on
off
Background illumination off
Auto
Backlighting automatically switched off 3
seconds after achieving stable weighing
value. Changes in weight or pressing of keys
will automatically result in backlight switching
on again.
Timer OFF
Automatic switch-off function in
stand-by mode
(see chap. 11.9)
Disable
AUTO-OFF switched off
2 minutes
AUTO-OFF after 2 minutes without changing
the weight
5 minutes
AUTO-OFF after 5 minutes without changing
the weight
15 minutes
AUTO-OFF after 15 minutes without
changing the weight
Time and date
(see chap. 11.10
Setting date and time
Language
(see chap. 11.11)
English / Deutsch
Language operator guide
English
Italiano
Calibration mode
(see chap. 11.12)
External calibration
Adjustment with external weight
Auto calibration
Automatic adjustment with internal
weight
Internal calibration
Adjustment with internal weight
after pressing CAL
Technical calib.
Overwriting the internal adjustment weight
Calibration info
(see chap. 11.13)
Record output of the last
adjustment
Return
Back to weighing mode